- Info CenterHalogen SensiCLĒNE™ The MP5™ was created with stable calibrations and low maintenance as priority requirements. To that end,...Real Talk about "Continuous Monitoring" Chlorine Analyzer claims Some manufacturers of online chlorine analyzers are making bold claims...
- Industries & ApplicationsMaritime & Aviation
Potable Water
Airport & AirPLane Monitoring
Inflow Source Monitoring
Tanker Monitoring
INDUSTRIALCommercial Buildings
Industrial Processing
Applications - ProductsHalogen SensiCLĒNE™ The MP5™ was created with stable calibrations and...Real Talk about "Continuous Monitoring" Chlorine Analyzer claims Some manufacturers...
- Contact
HSI Chlorine Analyzer - FAQ
Engineers and Chiefs Want to Know:
We often get the same questions about our chlorine analyzers due to their unique nature and claims we make about their accuracy and low maintenance requirements. Here are just a few common questions we get and their answers (click the header to find the answer):
The electrodes on the amperometric chlorine sensor should last the life of the system. No maintenance is usually necessary. In the event of abuse, the sensor may be reconditioned by simply repolishing it.
The sensor measures five parameters: Free Chlorine, pH, Temperature, Conductivity/Salinity, and ORP (oxidation reduction potential) every 50 seconds.
We recommend a calibration check every 60 days and pH calibration check every year. In addition, an annual replacement of the wear parts (impeller, o-rings, cleaning beads, and screen) and a bi-annual replacement of the pH sensor.
Yes. It will measure accurately from 200 µS to fully marine seawater, typically 45 mS (34 PSU). It will likewise measure accurately in waters with a pH of 6.5 to 9.
While the pH of seawater does not change much, in brackish or freshwater, the pH can change drastically from 6.5 to 9.0.
Halogen’s ORP measurement technique is different than a conventional dedicated ORP unit. It could vary by as much as 50 mV but should provide qualitative readings comparable to a dedicated system. It is not subject to poisoning and will respond faster to changes in some cases.
Halogen uses a unique pH cartridge in its sensor system. The pH cartridge can withstand 10 bar to -0.7 bar pressures and rapidly wets if allowed to dry out. Once reimmersed in water, the sensor’s accuracy restores in less than 3 minutes. Typically it does not require recalibration for a year or more, and the cartridge only needs replacing every two years.
Yes. AATEX/IECEx and the Non-Hazardous versions are both Type Approved by the Korean Registry of Shipping (KR)
Environmental Benefits & Testing
Our chlorine sensors have been designed and engineered with the environment in mind.
Our products are in compliance with:
- LR Rule
- DNV 2.4
- Tests included
- Vibration
- Susceptibility
- EMI
- Temp & Humidity Cycling
Still Have Questions? We Would Love to Hear From You...
See additional information about the benefits of the Halogen Systems amperometric chlorine sensor technology or just reach out to us directly and let us answer any questions you might have.