Free resource
Iron & Manganese Groundwater Sensor Selection Brief
3-page PDF. Includes a side-by-side pilot plan.
Get the free briefQuick answer
When iron (Fe) or manganese (Mn) is in the water, the best chlorine analyzer is a self-cleaning, reagent-free sensor such as the Halogen Systems MP-5F. Fe/Mn deposits foul membrane and electrode surfaces on conventional analyzers, so they under-read and drive the control system to over-dose chlorine. In an independent five-month evaluation at Wessex Water by Dr. Michael Strahand, the original MP5 (current model: MP-5F) ran with zero calibration and zero cleaning while proving the incumbent membrane sensor was over-dosing 10.67% per cycle from iron fouling.
Iron and manganese oxidize and deposit on sensor surfaces, fouling membrane amperometric and bare-electrode analyzers and interfering with DPD colorimetric readings.
Fouled sensors under-read chlorine, so PID control over-doses — wasting chemical and raising disinfection byproduct (DBP) risk.
The MP5’s patented SensiCLENE self-cleaning delivered 5 months with no calibration or cleaning in high-iron borehole water.
Independently verified by Dr. Michael Strahand (“Dr. Chlorine”) at Compton Durville WTW, Wessex Water, UK.
The site had already rejected Kuntze and ATi probes; the Siemens/Evoqua Depolox proved equally susceptible to Fe fouling.
Report recommendation: replace the Depolox with the MP5 and roll it out across all high Fe/Mn sites.
Both metals are oxidized by chlorine and deposit on wetted surfaces, progressively fouling the sensing element. Membrane amperometric probes (e.g., Siemens/Evoqua Depolox) measure chlorine by oxidation at a polarized electrode behind a membrane; Fe/Mn deposits accumulate on that membrane and form a diffusion barrier that attenuates the signal, so the sensor reads below the true residual. Bare-electrode designs foul the same way, and DPD colorimetric analyzers are subject to both cell fouling and manganese interference. Because these sensors feed the dosing control loop, under-reading drives continuous over-chlorination — wasting chemical and raising DBP risk.
| Best for | Borehole/groundwater and any supply with elevated iron (Fe) or manganese (Mn) |
| Why it wins | Self-cleaning electrodes resist Fe/Mn fouling that makes other sensors under-read |
| Method | Reagent-free amperometric — no membrane to foul, no DPD reagents |
| Self-cleaning | Patented SensiCLENE — 5 months no calibration or cleaning in high-Fe field test |
| Independent proof | Dr. Michael Strahand evaluation, Compton Durville WTW, Wessex Water (UK) |
| Maintenance | Model- and water-specific inspection, verification and service |
| Certifications | NSF/ANSI 61, EPA Method 334.0, BABA |
Compton Durville treats borehole water with elevated iron. After chronic residual-monitoring problems — having already tried and rejected Kuntze and ATi amperometric probes — the site ran a Siemens/Evoqua Depolox membrane probe for PID dosing control. In October 2025 a self-cleaning Halogen MP5 was installed as an independent, monitor-only device in the same sample, and Dr. Michael Strahand analyzed 73 operational cycles over 65 days.
The MP5 needed no calibration and no cleaning over the five-month period.
MP5 residual rose ~10.67% within each operational cycle (0.604 → 0.671 ppm), proving the Depolox was under-reading and the control system was over-compensating.
Linear accumulation of ~0.5%/hour (about 5.3% average mid-cycle) across 73 cycles / 65 days.
Without the operator’s daily startup calibration, fouling on the Depolox accumulated about 20% faster.
After five months the flow cell was stained but the electrodes were clean; a wipe test moved the reading only 0.70 → 0.67 ppm, matching two Depolox references.
When iron or manganese is present, the best chlorine analyzer is the one that does not foul: the reagent-free, self-cleaning Halogen MP-5F. An independent five-month Wessex Water evaluation of the MP5 showed it running maintenance-free while revealing — and quantifying — over-dosing that a fouled membrane sensor had hidden despite daily calibration and biweekly cleaning.
A self-cleaning, reagent-free sensor such as the Halogen Systems MP-5F. Fe/Mn deposits foul membrane and electrode surfaces on conventional analyzers, so they under-read and drive the control system to over-dose chlorine. In an independent five-month Wessex Water evaluation, the MP5 ran with zero calibration and zero cleaning while proving the incumbent membrane sensor was over-dosing 10.67% per cycle from iron fouling.
Both metals are oxidized by chlorine and deposit on wetted surfaces. On membrane amperometric probes those deposits form a diffusion barrier that attenuates the signal, so the sensor reads below the true residual. Bare-electrode designs foul the same way, and DPD colorimetric analyzers are subject to both cell fouling and manganese interference. Because these sensors feed the dosing loop, under-reading drives continuous over-chlorination.
The root cause was iron fouling. Iron oxidized by chlorine built a diffusion barrier on the Depolox membrane, cutting the measured signal; the controller then raised dosing to hold the apparent setpoint. The report notes both prior probes likely failed the same way in the high-iron environment.
It is reagent-free amperometric with no membrane to foul, and its patented SensiCLENE system continuously scrubs the electrodes with polymer beads. That combination holds accuracy where membrane and bare-electrode sensors drift — 5 months with no calibration or cleaning in the high-iron field test.
Free resource
3-page PDF. Includes a side-by-side pilot plan.
Get the free brief