Halogen-amperometrische Chlorsensoren

The Halogen MP5™

Übersicht

A breakthrough for drinking water and wastewater monitoring, The Halogen MP5™ Multiple Parameter Sensor tests 5 parameters using a 3 electrode amperometric chlorine measurement. The MP5™ is unaffected by flow or pressure, is self-cleaning, automatically compensates for pH, and is certified for NSF-61 use. The MP5™ utilizes Halogen’s patented RapidResponse ORP™ technology, a specialized implementation of ORP measurement that will report more consistent numbers thanks to a high resistance to poisoning.

The MP5™ requires little to no maintenance, can be installed in a flow cell, directly in a pipe, or immersed in a tank, channel, or basin. Simple to install and inexpensive to operate, the Halogen MP5™ is unlike any other sensor on the market today.

Einzigartige Merkmale

  • 5 Parameter in einem einzigen Online-Nahzeit-Analysator
    • Freies Chlor
    • pH-Wert
    • Temperatur
    • Leitfähigkeit
    • Schnelle Reaktion ORP™*
  • Selbstreinigend & wartungsarm
  • Keine Membranen oder Elektrolyte
  • NSF-61: Kein Abfallstrom erforderlich
  • Werkskalibriert, minimale Drift
  • Flexible deployment types: directly into a wet tap valve, pvc tees, side stream, or submerged
  • The system is equipped with 4 20mA and MODBUS RTU. Other communication protocols are available upon request.
  • 24 VDC oder 90 bis 240 VAC Betrieb

Benefits of the MP5™

Durchflussunabhängige Messung 

Der Sensor ist unbeeinflusst von Änderungen der Strömungsgeschwindigkeit von 0 bis 4 m/s. Eine integrierte Pumpe, die einen langlebigen, bürstenlosen Motor verwendet, liefert eine feste Geschwindigkeit über die Elektroden und schafft so Unabhängigkeit von der Strömung in jeder Installation. Die hohe Strömungsgeschwindigkeit über die Elektroden verbessert die Empfindlichkeit und senkt das Signal-Rausch-Verhältnis.

SensiCLĒNE™ Patentierte Selbstreinigungs-Technologie

Eine patentierte elektrochemische Reinigungsmethode namens SensiCLĒNE™widersteht der Polarisierung und Adsorption von organischen Stoffen. Die Reinigungsperlen reiben den Sensor kontinuierlich ab. Dadurch bleiben die Elektrodenoberflächen, einschließlich des pH-Sensors, frei von Ablagerungen, Kesselsteinbildung und Biofouling. Solide Metallelektroden halten die gesamte Lebensdauer des Systems. 

Diese Eigenschaften ermöglichen sehr lange Intervalle zwischen den Kalibrierungen.  

Kein Abfallstrom

Da der Sensor für die Verwendung nach NSF-61 zugelassen ist, kann er als Side-Stream-Einheit, direkt in einer Rohrleitung oder in einem Tank ohne erforderlichen Abwasserstrom installiert werden. 

Einfach zu überwachen

Ein optionales lokales Display ist für die Anzeige von Parametern, die Fehlersuche und die Kalibrierung von bis zu 5 separaten MP5™-Geräten erhältlich.

Servicefreundlichkeit

A Wet Tap installation method enables direct insertion into a distribution pipe. Sidestream, Immersion, and PVC Tee deployments are available. Whatever deplyment, the highly integrated design reduces the electronics cost and size, while increasing reliability.

Modelle und Bereitstellungsarten

MP5™ H1 Wasserhahn (H1WT)

Halogen MP5™ HotTap/WetTap-Sensoren

Das Wet Tap (oder Hot Tap)-Gehäuse kann als einziehbarer, direkter Rohreinschub für die Verwendung nach NSF-61 eingesetzt werden. Optionen für eine leckagefreie Aufrollvorrichtung sind verfügbar. 

MP5™ H1 Mittelflansch (H1MF-p)

Halogen MP5™-Sensoren mit PVC-Gehäuse

Das Mittelflanschgehäuse (H1MF-P) und die dazugehörige 2″-Adapterhülse und der Kragensatz ermöglichen einen einfachen Einsatz des NSF-61-PVC-T-Stücks im Seitenstrom.

H1 Niedrig FLANGE (H1LF-P)

Halogen MP5™ Sensoren mit zwei Arten von Durchflusszellengehäusen.

Das Modell mit niedrigem Flansch (H1LF-P) kann als Ersatz für bestehende, auf einer Steuertafel montierte Geräte oder als NSF-61-Tauchanalysator in einem Tank verwendet werden.

MP5™ Sensors & Specifications

Comparing MP5™ Deployment Specs

Specification
Wet Tap (D-H1WT-P)
PVC Tee (D-H1MF-P)
Sidestream/Immersion (D-H1LF-P)
Parameters (5)
  1. Free Chlorine
  2. pH-Wert
  3. Temperatur
  4. Leitfähigkeit
  5. Rapid ORP™*
  1. Free Chlorine
  2. pH-Wert
  3. Temperatur
  4. Leitfähigkeit
  5. Rapid ORP™*
  1. Free Chlorine
  2. pH-Wert
  3. Temperatur
  4. Leitfähigkeit
  5. Rapid ORP™*
Method
Reagentless, Self-cleaning, 3 electrode, no membrane or electrolyte
Reagentless, Self-cleaning, 3 electrode, no membrane or electrolyte
Reagentless, Self-cleaning, 3 electrode, no membrane or electrolyte
Measurement Range
0-10 ppm
0-10 ppm
0-10 ppm
LOD
30 ppb (0.03 ppm)
30 ppb (0.03 ppm)
30 ppb (0.03 ppm)
LOQ
90 ppb (.09 ppm)
90 ppb (.09 ppm)
90 ppb (.09 ppm)
Resolution
0.001 ppm (1 ppb)
0.001 ppm (1 ppb)
0.001 ppm (1 ppb)
Accuracy (FCl)
HSI’s analyzers can be deployed in environments with variable pH and temperature. For specific environments, please contact us.
Turbidity in Sample
(no effect)
3000 ppm (Arizona Test Dust Fine, 50 micron size)
3000 ppm (Arizona Test Dust Fine, 50 micron size)
3000 ppm (Arizona Test Dust Fine, 50 micron size)
Flow Spec
0 to 4.5 M/Sec (14.7 F/Sec)
0 to 4.5 M/Sec (14.7 F/Sec)
Bare: 0 to 4.5 M/Sec (14.7 F/Sec)
Flowcell: 1 to 30 LPH (0.3 to 8 GPH)
Water Use
None (NSF-61 models)
None (NSF-61 models)
None (NSF-61 models)
Pressure Range
-.07 to 10 bar (-10.15 to 147 psi)
-.07 to 10 bar (-10.15 to 147 psi)
-.07 to 10 bar (-10.15 to 147 psi)
Maintenance Schedule
1 yr replace SensiCLĒNE™ wear-parts.
Calibration at 3 to 6 mo or as necessary
1 yr replace SensiCLĒNE™ wear-parts.
Calibration at 3 to 6 mo or as necessary
1 yr replace SensiCLĒNE™ wear-parts.
Calibration at 3 to 6 mo or as necessary
Calibration
Factory Calibrated
Factory Calibrated
Factory Calibrated
Calibration Stability
6mo (typical)
6mo (typical)
6mo (typical)
Interferances
ClO2, Ozone
ClO2, Ozone
ClO2, Ozone
* RapidORP™ is HSI’s proprietary indicator for contamination. HSI’s measurement technique and self-cleaning system eliminate poisoning issues, providing a more reliable reading. HSI’s RapidORP™ measurement may have a slight offset from other sensors, it provides a reliable, qualitative measurement that overcomes many of the ORP problems. See the explanatory ORP vs Amperometric Sensors documentation.

Handbücher für Trinkwasser- und Abwasseranwendungen

Installationshandbuch

D20 Display/Controller & Sensor for use in Drinking Water and Wastewater applications.

Betriebshandbuch

MP5 Online Chlorine Analyzer Manual (2024.06.12-v4.02) for use in Drinking Water and Wastewater applications.

Installationshandbuch

Installationshandbuch für MP5™ D-H1WT-P Chloranalysator im Wet Tap Einsatz mit Entnahmewerkzeug  

Installationshandbuch

Laden Sie das Handbuch zum Austausch von Verschleißteilen für die SensiCLĒNE™-Selbstreinigungstechnologie in MP5™-Analysatoren herunter

MP5™ LÖSUNGEN UND KITS

Kit with flow-cell deployment (MKIT) and display controller along with a wall mounting plate,  provides a drop-in upfit sensor replacement.

Mehr Informationen

Batteriestrom und Ferntelemetrie für einen MP5™-Einsatz. Für die Überwachung des äußeren Netzes, Sackgassen oder andere Remote-Implementierungen.

Mehr Informationen

Nutzen Sie die Zuverlässigkeit und den geringen Wartungsaufwand des MP5™ für die Tankfernüberwachung mit Batteriebetrieb und Ferntelemetrie.

Mehr Informationen

Das ausziehbare Entnahmeset für den MP5™ Wet-Tap-Analysator (H1WT-P) in Wasserleitungen ermöglicht eine Wartung ohne Unterbrechung des Durchflusses.

Mehr Informationen

Dieser Einsatz erfolgt mit einem wasserdichten MP5™ mit niedrigem Flansch (H1LF-P), einem gesourcten Ausleger und Gurtzeug oder kann einfach in einem Tank oder Trog platziert werden.

Mehr Informationen

Kombinieren Sie einen Mittelflansch MP5™ (H1MF-P) mit einer Adapterhülse in PVC- oder Edelstahlgröße für den Einsatz in einem 2″-PVC-T-Stück.

Mehr Informationen

Passt zum MP5™ mit niedrigem Flansch (H1LF-P) und passt sich je nach bestelltem Durchflusszellentyp an die Verwendung von Seitenstrom und Steuerplatine an. 

Mehr Informationen

Both 24V DC (DC20-L) and 120-240V AC (DC20-H) versions feature (4) 4-20 mA analog outputs, and an M-12 Connector, 3 relays, SD card. 

Mehr Informationen